What is an NIR Calibration

NIR (Near InfraRed spectroscopy) is a versatile technology that is widely used to provide reliable and accurate analysis for a wide range of sample types.  The ability to provide accurate results relies on two factors:

  • A high quality NIR spectrum obtained from a reflectance or transmission analysis of the sample.
  • A robust calibration correlating the NIR spectrum to the component of interest. 
What is an NIR Calibration?

As decribed in an earlier blog, NIR is an analytical technique that is based on measuring differences in how NIR energy is transmitted or reflected off of a sample, depending on the chemical composition of the sample. Thus NIR is a secondary technique, meaning that a primary reference method is required to create an NIR calibration.  The calibration itself is a mathematical correlation between the raw NIR data from samples and the chemical consituent or property of interest. 

How is an NIR Calibration Made?

Calibration Sample SpectraThe first task for developing an NIR calibration is to collect samples. The samples should be representative of the future unknown samples to be measured in all areas of potential variability including consituent range(s), origin, seasonal variation etc. 

For full spectrum instruments such as PDA, FT and scanning monochromators, a sufficient number of samples is required to properly account for the variability of the samples across the many datapoints collected by these instruments. A minimum number of samples to get a provisional calibration is normally 50 samples, but good robust calibrations are usually over 100 samples depending on the product type, and some can be based on thousands of samples.

A first collection of selected samples is then sent for reference analysis.  This might be Dumas or Kjeldahl for protein, Soxhlet extraction for fats and oils, or any other approved reference method for the consituent(s) of interest.  It is important that the reference data be of high quality and with minimal error are the accuracy and performance of the NIR calibration is limited by the reference error.

Once the reference data are obtained, they are added to raw sample spectra and these data are regressed against each other, often using a PLS regression.  The output is a linear equation that can be applied to future unknown samples to predict the constituents of interest.  A diagram is below:

NIR Calibration Procedure Diagram

Check this box if you would like to receive updates on future product releases, events and company news from Unity Scientific
* Indicated Required Field
Check this box if you would like to receive updates on future product releases, events and company news from Unity Scientific
* Indicated Required Field
Check this box if you would like to receive updates on future product releases, events and company news from Unity Scientific
* Indicated Required Field